| Model Number |
FIBERTAK DR LBRLTAPE WH/BL & TT WH/BLK |
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
12/07/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2021 it was reported by a sales representative via email that an ar-3651tt, fibertak dr lbrltape, was being used in an rcr procedure.During insertion, the fork tip of the anchor broke off within the patient's joint and was unable to be retrieved.The case was completed with a second anchor with the same pn without further incident.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint confirmed.Visual evaluation confirms that one of the fork tips, at the distal end has broken off.Broken piece and suture were not returned for evaluation.Device functioning was not performed due to the broken condition of the device.The cause of this event is undetermined; however, a most likely cause of this event is the application of prying/leveraging forces to the device during use.
|
| |
|
Search Alerts/Recalls
|