H10: h3, h6: the reported device was received for evaluation.There was a relationship found between the device and the reported event.A visual inspection found no issues.A functional evaluation revealed both p1 and p2 transducers failed the high pressure tests.Both were reading a lower pressure than the actual input.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The complaint was confirmed, and the root cause was associated with an electrical component failure.Factors which could have contributed to the reported event include a defective transducer.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
|