• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. IV3000 1 HAND 10X12CM CTN 50; TAPE AND BANDAGE, ADHESIVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. IV3000 1 HAND 10X12CM CTN 50; TAPE AND BANDAGE, ADHESIVE Back to Search Results
Model Number 66024008
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Rash (2033); Blister (4537)
Date of Event 08/25/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, dressing was used to fix the picc tube, 5 days later the patient experienced scattered rashes and blisters with pruritus on the skin where the dressing of a iv3000 1 hand 10x12cm ctn 5 was applied.This adverse event was treated by oral antiallergic drugs, external application of trimetazidine cream on (b)(6) 2021.It is unknown current health status of patient.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Additional Manufacturer Narrative
H3, h6: the batch record was reviewed and it could be confirmed that the products had progressed through a satisfactory release process which involved testing the product against the requirements of the finished product specification.There was also no issue identified during the manufacturing process that could have caused or contributed to the complaint problem.A complaint history review revealed a small number of similar instances in the last three years.There is nothing to indicate that this is outside of acceptable rates of occurrence.The device was used for treatment and was not returned for analysis.We have not been able to confirm a relationship between the event and the device or identify a definitive root cause.It was reported that 5 days after the dressing was applied the patient experienced scattered rashes and blisters.A clinical assessment determined that without clinically relevant patient-specific supporting documentation, a thorough medical investigation cannot be performed.The risk files mitigate the reported issue with no updates required.Ifu contains comprehensive instructions on skin preparation and application of dressing and also possible adverse reactions and wound suitability.A probable root cause for this issue is incorrect application of dressings or infrequent dressing changes.The users of the reported product are advised to consult the ifu, to prevent future occurrences of the reported issue.This guide provides comprehensive instructions of the operation, use and limitations of the device, including application and removal of dressings and skin preparation prior to use.This investigation is now complete, with no corrective actions required.No further actions by smith and nephew are deemed necessary at this stage.However, we will continue to monitor for any adverse trends relating to this product range.Smith and nephew acknowledge customer concern and are continually investigating ways to develop and improve our products.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IV3000 1 HAND 10X12CM CTN 50
Common Device Name
TAPE AND BANDAGE, ADHESIVE
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
Manufacturer (Section G)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK   HU3 2BN
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13090553
Report Number8043484-2021-02063
Device Sequence Number12092291
Product Code KGX
UDI-Device Identifier05000223417406
UDI-Public05000223417406
Combination Product (Y/N)N
Initial Reporter CountryCH
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/01/2023
Device Model Number66024008
Device Catalogue Number4008
Device Lot Number2040
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/06/2021
Supplement Date Received by Manufacturer02/25/2022
Initial Report FDA Received Date12/27/2021
Supplement Report FDA Received Date02/28/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/28/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age45 YR
Patient SexFemale
-
-