| Model Number |
PCB00 |
| Medical Device Problem Codes |
Difficult to Fold, Unfold or Collapse (1254); Inaccurate Delivery (2339); Failure to Advance (2524)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event: the exact date of event is unknown, not provided, but the best estimate date is october/ november 2021.Implant date: unknown/ not provided.Explant date: lens remains implanted, therefore not explanted.Telephone number: (b)(6).The lens was not returned for evaluation as it remains implanted.Therefore, a failure analysis of the complaint device could not be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that after implantation according to the dfu (directions for use), the last push was given and the trailing haptic unfolded and stayed in the cartridge.The haptic was not folded on the optic of the lens and the doctors got it out of the cartridge without damage, requiring a secondary tool to free the lens.There was a delay in procedure; however, no patient issues reported and no medical or surgical interventions required.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Section d9 - device available for evaluation? yes section d9 - returned to manufacturer? yes section d9 - date returned to manufacturer: 10th january 2022 section h3 - device evaluated by manufacturer? yes device evaluation: visual inspection under magnification revealed viscoelastic residue inside of the cartridge.The handpiece was disassembled and the assembly was inspected, presenting with no issues.No complaint lens was received as part of this return, therefore, no evaluation could be performed on the lens.The complaint issues could not be confirmed, and no product deficiency could be identified.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|