• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL INC. ARROW MULTI-LUMEN PI CVC KIT: 3-L 7 FR X 8" (2; CATHETER,INTRAVASCULAR,THERAPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL INC. ARROW MULTI-LUMEN PI CVC KIT: 3-L 7 FR X 8" (2; CATHETER,INTRAVASCULAR,THERAPE Back to Search Results
Model Number IPN037072
Medical Device Problem Code Deformation Due to Compressive Stress (2889)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/08/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
Customer reports guidewire kinks and another kit had to be open for a new wire.Also "had to move to a second site a few times because of it".No patient harm reported.There was a delay in care related to extended procedure time and it was reported there was an increased risk of infection related to multiple site access.
 
Event or Problem Description
Customer reports guidewire kinks and another kit had to be open for a new wire.Also "had to move to a second site a few times because of it".No patient harm reported.There was a delay in care related to extended procedure time and it was reported there was an increased risk of infection related to multiple site access.
 
Additional Manufacturer Narrative
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW MULTI-LUMEN PI CVC KIT: 3-L 7 FR X 8" (2
Common Device Name
CATHETER,INTRAVASCULAR,THERAPE
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
Manufacturer (Section G)
ARROW INTERNATIONAL DE MEXICO S.A. DE C.V.
ave. washington 3701
colonia panamericana, chihuahua
chihuahua 31200
MX   31200
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key13094005
Report Number9680794-2021-00687
Device Sequence Number11306554
Product Code FOZ
UDI-Device Identifier30801902119644
UDI-Public30801902119644
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K071538
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2023
Device Model NumberIPN037072
Device Catalogue NumberASK-45703-AKP1A
Device Lot Number13F21H1178
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/08/2021
Supplement Date Received by Manufacturer01/21/2022
Initial Report FDA Received Date12/28/2021
Supplement Report FDA Received Date01/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/08/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-