| Model Number |
5584180 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Fracture (1260); Deformation Due to Compressive Stress (2889)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/01/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As the lot number for the device was provided, a review of the device history record is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.(expiry date: 05/2026).
|
| |
|
Event or Problem Description
|
|
It was reported that during a port placement procedure in the subclavian vein, the pink hub attached to the introducer needle was allegedly fractured.It was further reported that the needle allegedly leaked on the outside which in turn delayed the treatment.The procedure was completed by using another device.There was no reported patient injury.
|
| |
|
Event or Problem Description
|
|
It was reported that during a port placement procedure in the subclavian vein, the pink hub attached to the introducer needle was allegedly fractured.It was further reported that the needle allegedly leaked on the outside which in turn delayed the treatment.The procedure was completed by using another device.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one introducer needle was returned for evaluation.Gross visual, microscopic and functional testing were performed.The investigation is confirmed for the identified bent, reported needle fracture and leak issue as bend was noted on the needle and multiple cracks were observed on the proximal hub of the introducer needle and leaks were noted from the cracks on the hub upon infusion of the introducer needle.The definitive root cause could not be determined based upon available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 05/2026).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
|
| |
|