• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SYNERGY; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SYNERGY; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number 10619
Device Problems Failure to Advance (2524); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/09/2021
Event Type  malfunction  
Event Description
It was reported that stent damage occurred.Vascular access was obtained via the right artery.The 92% stenosed, 16x5mm, de novo, eccentric target lesion with a significant bend of >=90 degrees was located in the severely tortuous and severely calcified right coronary artery.A 4.00 x 20mm synergy ii drug-eluting stent was advanced but failed to cross the lesion.The device was removed and it was found that the tip of the stent itself was deformed.The physician decided to complete the procedure by plain old balloon angioplasty (nc emerge 4.5/15).There were no patient complications reported and the patient status was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY
Type of Device
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13097189
MDR Text Key282911822
Report Number2134265-2021-16252
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)Y
Reporter Country CodeTW
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Nurse
Type of Report Initial
Report Date 12/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/15/2022
Device Model Number10619
Device Catalogue Number10619
Device Lot Number0025756499
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/09/2021
Initial Date FDA Received12/28/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/15/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age48 YR
Patient SexMale
-
-