| Model Number |
2076-5040 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/10/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The angiosculpt device was used to treat a severely calcified mid iliac.During the 4th inflation, the balloon ruptured at 14 atm.The procedure was completed with a non-angiosculpt device.No patient injury reported.This product problem is being submitted per fda request because the balloon ruptured at rbp.
|
| |
|
Additional Manufacturer Narrative
|
|
The patient's dob or age at time of event, gender, weight, ethnicity, and race are unknown.This information was not available from the facility.Patient information regarding relevant tests/ laboratory data or medical history are unknown.This information was not available from the facility.Report source: foreign- (b)(6).The angiosculpt device was discarded by the facility, thus no returned product investigation was performed.
|
| |
|
Additional Manufacturer Narrative
|
|
Block b5: updated from "this product problem is being submitted per fda request because the balloon ruptured at rbp" to "this product problem is being submitted conservatively because the balloon ruptured at rbp.".
|
| |
|
Event or Problem Description
|
|
This product problem is being submitted conservatively because the balloon ruptured at rbp.
|
| |
|