Model Number 72200616 |
Device Problems
Overheating of Device (1437); Protective Measures Problem (3015)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/06/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Internal complaint reference: (b)(4).
|
|
Event Description
|
It was reported that during a knee scope procedure, the cord of the mdu was overheating and an error code appear on the dii console.The procedure was successfully completed with no delay using a back-up device.No patient complications were reported.
|
|
Manufacturer Narrative
|
Internal complaint reference (b)(4).The reported device was received for evaluation.There was a relationship found between the returned device and the reported event a visual inspection found no issues.A functional evaluation revealed overheating of the housing while the mdu was running.No error message was noted.The motor/gearbox could not be removed from the housing for further assessment.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The complaint was confirmed, and the root cause was associated with a mechanical component failure.Factors which can contribute to overheating include a blade stall condition that will result in increased current draw from the control unit which will heat the motor and hand piece housing.This can be the result of gearbox corrosion caused by cleaning and sterilization methods and the chemicals involved.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.Internal complaint reference case-2021-00084638-1 h6: failure mode updated.B5: event description updated.
|
|
Event Description
|
It was reported that during a knee scope procedure, the cord of the mdu was overheating.The procedure was successfully completed with no delay using a back-up device.No patient complications were reported.
|
|
Search Alerts/Recalls
|