C.R. BARD, INC. (COVINGTON) -1018233 BARD® INLAY® URETERAL STENT Back to Search Results
Model Number 778424
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/29/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the stent was found in the abdomen outside the ureter.There was a pigtail catheter adjacent to the ureter.This was external to the ureter and not in bowel.Clinical correlation was recommended and this has been placed in the interval.
 
Event or Problem Description
It was reported that the stent was found in the abdomen outside the ureter.There was a pigtail catheter adjacent to the ureter.This was external to the ureter and not in bowel.Clinical correlation was recommended and this has been placed in the interval.
 
Additional Manufacturer Narrative
The reported event is inconclusive because no sample was returned.A potential root cause for this event could be, "design qualification" or "design output".The device was used for treatment purposes, however, it is unknown if the device had met relevant specifications or contributed to the reported event.No manufacturing issues or non-conformances were noted in the dhr review that could have caused or contributed to the reported event.Based on the results of the investigation no additional action is required at this time.The device was not returned for evaluation.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "with any ureteral stent, migration is a possible complication which could require medical intervention for removal.Selection of too short a stent may result in migration." "determine the proper stent length for the patient.This is generally calculated from the baseline pyelogram.Accurate measurements will optimize drainage effi ciency and patient comfort.Submerge stent in sterile water to activate the coating." the device was not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® INLAY® URETERAL STENT
Common Device Name
URETERAL STENT
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key13102512
Report Number1018233-2021-08497
Device Sequence Number14301788
Product Code FAD
UDI-Device Identifier00801741014758
UDI-Public(01)00801741014758
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
K983498
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/20/2022
Device Model Number778424
Device Catalogue Number778424
Device Lot NumberNGBW3961
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/15/2021
Supplement Date Received by Manufacturer04/28/2022
Initial Report FDA Received Date12/28/2021
Supplement Report FDA Received Date05/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/11/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;