• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 ALYTE¿ Y-MESH GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 ALYTE¿ Y-MESH GRAFT Back to Search Results
Model Number Y500
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pain (1994); Dyspareunia (4505)
Date of Event 11/30/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the alyte y-mesh graft exposure at vaginal vault.The patient underwent laparoscopic sacrocolpopexy and cystoscopy and experienced dyspareunia and pain in the vagina.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the alyte y-mesh graft exposure at vaginal vault.The patient underwent laparoscopic sacrocolpopexy and cystoscopy and experienced dyspareunia and pain in the vagina.
 
Additional Manufacturer Narrative
The reported event was inconclusive because no sample was returned.It is unknown whether the device had met relevant specifications.The product was used for treatment purposes.It was unknown whether the product had caused the reported failure.A potential root cause for this failure could be "graft improperly tensioned during placement (physician error)".The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "precautions: based on physician experience and training, a thorough assessment of each patient should be made to determine their suitability for a synthetic mesh procedure.Additional consideration should be given to the use of alyte® y-mesh graft in patients with a compromised immune system, any condition that would compromise healing, or any patient with a history of prior abdominal or pelvic surgeries.Consideration should also be given to the ability of the patient to tolerate the surgical procedure.Accepted surgical practice and precautions must be followed for the management of infected or contaminated wounds.Postoperative bleeding may occur in some patients and must be controlled prior to patient release.Sutures should not be placed in the mesh edge.Sutures should be placed a minimum of 1cm from any mesh edge.Inadequate suturing of the graft material to the pelvic tissues may lead to failure of the repair, additional complications and recurrence of prolapse.Check the integrity of the packaging before use.Do not use the mesh if the packaging is opened or damaged.As for any implantable material, it is recommended to open the package at the time of implantation.The alyte® y-mesh graft is intended as a single-use device.Do not re-sterilize any portion of the alyte® y-mesh graft.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient.Post-operatively the patient should be advised to refrain from heavy lifting, exercise (e.G.Cycling, jogging) and/or intercourse until the physician determines it is suitable for the patient to return to normal activities.Adverse events: complications associated with the proper implantation of the alyte® y-mesh graft may include, but are not limited to those typically associated with surgically implantable materials including: postoperative hematoma, seroma, abscess or fistula formation, or scarring which may occur following the implant procedure.Urinary retention, bladder outlet obstruction and other voiding and defecatory dysfunctions.These conditions may be associated with over-correction/too much tension placed on the implant.Perforations or lacerations of vessels, nerves, bladder, bowel, urethra, rectum, or any viscera, which may occur during the implantation procedure.Irritation at the operative wound site which may elicit a foreign body response that leads to wound dehiscence, inflammation and/or infection.Extrusion through vaginal epithelium or erosion into surrounding viscera and/or mucosa.Inflammation, sensitization, pain, dyspareunia, scarification, contraction, device migration and failure of the procedure resulting in recurrence of vaginal wall prolapse.Urinary incontinence (stress and urge)." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that the alyte y-mesh graft exposure at vaginal vault.The patient underwent laparoscopic sacrocolpopexy and cystoscopy and experienced dyspareunia and pain in the vagina.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALYTE¿ Y-MESH GRAFT
Common Device Name
Y-MESH GRAFT
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key13102533
Report Number1018233-2021-08501
Device Sequence Number19710656
Product Code OTO
UDI-Device Identifier00801741016165
UDI-Public(01)00801741016165
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K101722
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/31/2017
Device Model NumberY500
Device Catalogue NumberY500
Device Lot NumberHUZA0679
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/15/2021
Supplement Date Received by Manufacturer02/24/2022
Initial Report FDA Received Date12/28/2021
Supplement Report FDA Received Date03/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/09/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-