| Model Number |
24-025-51-09 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Impaired Healing (2378)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Multiple mdr reports were filed for this patient.Mdr 9610905-2021-00123.
|
| |
|
Event or Problem Description
|
|
The physician reported that a sternal plate broke due to patient non-compliance.The plate was removed and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation was performed in the lab and there were no indications of material or manufacturing defects.The complaint percentage was calculated, and it is determined that the complaint percentage falls within the design risk limits adhered to at klm.During the investigation the product lot number identified was reviewed in the device history records.The dhr review showed no discrepancies or anomalies.The investigation results conclude that the root cause is patient related as no device issue was found.If further information is obtained that might add value to the contents of the investigation report, an additional follow-up report will be submitted.
|
| |
|