MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE® DRYSEAL FLEX INTRODUCER SHEATH; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE® DRYSEAL FLEX INTRODUCER SHEATH; INTRODUCER, CATHETER Back to Search Results
Catalog Number DSF2233
Device Problems Improper or Incorrect Procedure or Method (2017); Patient-Device Incompatibility (2682)
Patient Problem Vascular Dissection (3160)
Event Date 12/03/2021
Event Type  Injury  
Manufacturer Narrative
It should be noted the gore® dryseal flex introducer sheath instructions for use (ifu) state ¿adequate vessel access is required to introduce the sheath into the vasculature.Careful evaluation of vessel size, anatomy, tortuosity, and disease state (including calcification, plaque, and thrombus) is required to ensure successful sheath introduction and subsequent withdrawal.If vessel is not adequate for access, major bleeding, vessel damage, or serious injury to the patient, including death, may result.¿.
 
Event Description
On (b)(6) 2021, the patient underwent endovascular treatment of a thoracic aneurysm using gore® excluder® aaa endoprostheses.A 22fr gore® dryseal flex introducer sheath was also used as an accessory during the procedure.It was reported the 22fr sheath was inserted through the left common femoral artery for access.According to the report, the vessel was notably tortuous and narrow (measurement not provided), and resistance was felt while advancing the sheath.A dissection was reportedly identified at the insertion site, and surgical repair was required.A decision was made to access the right common femoral artery.The same 22fr sheath was inserted, and stent grafts were successfully advanced and implanted.Upon withdrawal of the 22fr sheath, a suspected dissection in the right common iliac artery was observed.An additional non-gore manufactured stent was implanted to treat the dissection on the right side.The patient tolerated the procedure.
 
Manufacturer Narrative
H6: added investigation conclusion code 50.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® DRYSEAL FLEX INTRODUCER SHEATH
Type of Device
INTRODUCER, CATHETER
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL PHOENIX 1 B/P
32360 n. north valley parkway
phoenix AZ 85085
Manufacturer Contact
damon jackson
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key13110008
MDR Text Key288184528
Report Number3007284313-2021-01730
Device Sequence Number1
Product Code DYB
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K160254
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberDSF2233
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/03/2021
Initial Date FDA Received12/29/2021
Supplement Dates Manufacturer Received06/20/2022
Supplement Dates FDA Received06/22/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age82 YR
Patient SexMale
-
-