Model Number 1504-10-125
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pain (1994); Osteolysis (2377); Joint Laxity (4526)
Date of Event 11/03/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Initial reporter occupation: lawyer.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Patient received a left attune knee to treat tricompartmental degenerative joint disease.The patella was resurfaced and competitor cement x 2 was utilized.There were no indicated intra-operative complications.Patient received a left knee revision to address instability, osteolysis in the lateral femoral condyle distally, pain, tibial tray migration, and tibial tray loosening at the cement to implant interface.The tibial tray, tibial insert, and femoral component were revised.The patella component was retained.The patient was revised with competitor products and cement.There were no indicated intra-operative complications.Doi: (b)(6) 2015, dor: (b)(6) 2020, left knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE PS FEM LT SZ 5 NAR CEM
Common Device Name
ATTUNE IMPLANT : KNEE FEMORAL
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI  
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key13110104
Report Number1818910-2021-28940
Device Sequence Number4717914
Product Code NJL
UDI-Device Identifier10603295041702
UDI-Public10603295041702
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2024
Device Model Number1504-10-125
Device Catalogue Number150410125
Device Lot Number7908532
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/16/2021
Supplement Date Received by Manufacturer02/07/2022
Initial Report FDA Received Date12/29/2021
Supplement Report FDA Received Date02/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/15/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE FB TIB BASE SZ 3 CEM; ATTUNE MEDIAL DOME PAT 35MM; ATTUNE PS FB INSRT SZ 5 6MM; ATTUNE PS FEM LT SZ 5 NAR CEM; PALACOS BONE CEMENT; PALACOS BONE CEMENT
Outcome Attributed to Adverse Event Required Intervention;
Patient Age56 YR
Patient SexFemale
-
-