The returned stone cone was analyzed, and a visual evaluation found the coil exposed and tangled.The green coating of the coil was damaged such that the inner coil wire is exposed.Additionally, the outer blue sheath had a slight tear at the distal end.Due to the defects, functional testing found the coil could not be successfully opened or closed.Therefore, the reported event is confirmed.Based on all available information, it is highly unlikely that the defects described were present upon opening of the packaging.It is most likely that they were caused using excessive force or manipulation during testing.The instructions for use (ifu) warns the user, prior to use, ensure that the coil is working properly by advancing the sheath over the coil to the positive stop and then retracting the sheath to open the coil.The sheath of the device should be straight during testing.Therefore, the most probable root cause is unintended use error caused or contributed to event.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.
|
It was reported to boston scientific corporation that a stone cone nitinol retrieval coil was to be used in the ureteral calculus ureteroscopic holmium laser lithotripsy procedure performed on (b)(6) 2021.During preparation, the stone cone was tested and found the device failed to form a coil.Another stone cone was used to complete the procedure.There were no patient complications as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.Note: this event has been deemed an mdr-reportable event based on the investigation results that the green coating of the coil was damaged and inner coil wire is exposed, and outer blue sheath had a slight tear at the distal end.
|