• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GII SPC REAMTHRU CR TR SZ 8 RT; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GII SPC REAMTHRU CR TR SZ 8 RT; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 71433358
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code Laceration(s) (1946)
Date of Event 12/06/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the gii spc reamthru cr tr sz 8 rt developed metal shard and cut the spd tech during cleaning process.The current state of health of the spd tech is unknown.
 
Additional Manufacturer Narrative
Internal complaint reference (b)(4).
 
Additional Manufacturer Narrative
H3, h6: the associated device was returned and evaluated.The manufacturing process of the device did not contribute to the reported event.A visual inspection of the returned device confirmed the stated failure mode.The device has several scratches and chips along its surface, rendering the device inoperative.The device shows signs of extensive use.The clinical/medical evaluation concluded that per case details, a metal chard developed and cut the employee (spd tech) during the cleaning process.No surgical case was in progress during this time.The current state of health status of the spd tech is unknown and additional clinically relevant information was not provided.Without the requested information, the root cause of the reported events cannot be definitively concluded and the employee impact beyond the reported cannot be determined.Therefore, no further clinical/medical assessment can be rendered at this time.A review of complaint history did not reveal similar events for the listed device.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII SPC REAMTHRU CR TR SZ 8 RT
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13111588
Report Number1020279-2021-08949
Device Sequence Number4589705
Product Code HRY
UDI-Device Identifier00885556032312
UDI-Public00885556032312
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K121393
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number71433358
Device Catalogue Number71433358
Device Lot Number11FM00655
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/22/2021
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/10/2021
Supplement Date Received by Manufacturer01/26/2022
Initial Report FDA Received Date12/29/2021
Supplement Report FDA Received Date01/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/01/2011
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-