H6: investigation summary: photos received for investigation.Upon visual inspection, an injector, connector, and needle is displayed.Unable to observe leakage on images received.A device history review was performed for reported lot 2102318, 2012504, no deviations or non-conformances were identified during the manufacturing process that could have contributed to this reported issue.Product undergoes inspections throughout the manufacturing process to ensure the quality and functionality of the device, including leakage and that all critical dimensions are within specification and there is no damage on the product.Five retained samples from the same lot are evaluated.Upon visual inspection of the samples requested, no defects can be observed.Further testing was conducted, no sign of leakage occurred.Functional testing was performed, penetrating the injector along with a sample syringe and mating components ten times, all results were found to be acceptable with no leaks identified.Based on the quality team's investigation, we are not able to identify a root cause related to our manufacturing process at this time.
|