Model Number ROB10016
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/10/2021
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference case-(b)(4).
 
Event Description
It was reported that during a cori assisted tka surgery, the welding of the real intelligence robotic drill guard broken at the part of suction tube attachment.The procedure was completed, without delay, using a s+n back-up device.Patient was not harmed beyond the reported problem.
 
Manufacturer Narrative
H3, h6: the real intelligence robotic drill guard, (japan) part number rob10016, lot1107083 used for treatment, was not returned for evaluation; thus, a visual and functional evaluation could not be performed, and a relationship between the reported event and the device could not be established.A review of manufacturing records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints found similar events.Although the reported problem was not confirmed a factor that may have contributed to the reported symptom may have been associated with an incomplete weld around the perimeter of the irrigation tube.A historical escalation event review was not completed.The product was not returned and no evidence was made available to link the complaint to an escalation event.The real intelligence cori for knee arthroplasty user manual provides guidelines for recovering to a fully manual procedure in the "recovery procedure guidelines, appendix d".Per our risk assessment the failure mode and associated risk have been anticipated within the risk file and the documented risk level is still adequate.Per complaint details from japan, it was reported that during a cori assisted tka surgery, while burring the femur the welding of the real intelligence robotic drill guard broke at the suction tube attachment.The procedure was completed without delay, using a s+n back-up device.Patient was not harmed beyond the reported problem.Although no further containment or corrective action is recommended or required at this time, all complaints are monitored and trended through post market surveillance activities.If the product associated with this event is returned or provided at a future date, this evaluation will be reopened for investigation.
 
Manufacturer Narrative
The manufacturer has identified that this event should be re-evaluated for mdr reporting.The detached component is of a size/shape that can be easily retrieved in its entirety and this malfunction has been resolved by using an available back-up device.Although this event may result in a short delay while the procedure is resumed and completed as intended, this malfunction has not caused or contributed, in the past, to a death or serious injury, nor has it been required a medical intervention to preclude permanent impairment of a body function or permanent damage to a body structure.This malfunction is unlikely to cause any death or serious injury if it were to recur, either.Therefore, this event is considered non-reportable pursuant to 21 c.F.R.803.If further details are provided confirming the occurrence of a reportable event, our files will be updated accordingly, and a further report submitted outlining both the event details and our investigations performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAL INTELLIGENCE ROBOTIC DRILL GUARD
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13113765
MDR Text Key282973765
Report Number3010266064-2021-00920
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00885556757352
UDI-Public00885556757352
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K193120
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 12/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/29/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberROB10016
Device Catalogue NumberROB10016
Device Lot Number1107083
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received12/22/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
ROB20000/CORI ROBOTICS USA
-
-