• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER TRAUMA KIEL LONG NAIL KIT R1.5, TI, RIGHT GAMMA3® Ø11X360MM X 125°; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER TRAUMA KIEL LONG NAIL KIT R1.5, TI, RIGHT GAMMA3® Ø11X360MM X 125°; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 3425-1360S
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 11/29/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device will not be returned.If additional information becomes available, it will be provided in a supplemental report.Device disposition is unknown.
 
Event or Problem Description
"client reported broken gamma 3 nail at 6 weeks post operative.Revision surgery needed to remove broken nail and replace with other implant".
 
Additional Manufacturer Narrative
The reported event was confirmed.Review of dhr, complaint history, capa databases and risk analysis did not identify any discrepancies.No nc/capa had been filed for the item in question.Potential nail breakage had been considered in the risk management file.Potential reasons for nail breakage are listed in the labelling.Physical investigation of the nail revealed it broke in a fatigue manner, originated by material damage at lateral in the posterior web of the proximal drill hole.Additionally, the nail was heavily loaded during the implantation period.Material and manufacturing deficiencies were not detected.X-rays were provided and were presented to a healthcare professional for review.His opinion was (in extracts): ¿very young patient¿seems to be in a very bad overall health status.¿if she is still having alcohol consumption, she may have been very careless with the weight bearing¿.Lag screw placed exactly through the laterally opened fracture.All of this may have contributed to the poor outcome.¿ with available information a deficiency of the device was not verified.The cause of the event was determined being patient related contributed by intra-operative material damage.¿ an unsuitable placement of the lag screw was concluded as root cause and being user related.In case further substantive information becomes available we reserve the right to re-open the investigation with root cause assessment.
 
Event or Problem Description
"client reported broken gamma 3 nail at 6 weeks post operative.Revision surgery needed to remove broken nail and replace with other implant".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LONG NAIL KIT R1.5, TI, RIGHT GAMMA3® Ø11X360MM X 125°
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
marilyne chaumont
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key13114424
Report Number0009610622-2021-00832
Device Sequence Number17317545
Product Code HSB
UDI-Device Identifier07613252273837
UDI-Public07613252273837
Combination Product (Y/N)N
Initial Reporter CountryBE
PMA/510(K) Number
K034002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date03/31/2026
Device Model Number3425-1360S
Device Catalogue Number34251360S
Device Lot NumberK0BBD18
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/29/2021
Supplement Date Received by Manufacturer02/28/2022
Initial Report FDA Received Date12/29/2021
Supplement Report FDA Received Date03/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/26/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age49 YR
Patient SexFemale
Patient Weight64 KG
-
-