• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. REMSTAR SE, W/ HEATED TUBE, SD; VENTILATOR, NON-CONTINUOUS (RESPIRATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. REMSTAR SE, W/ HEATED TUBE, SD; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number DS220TS
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/20/2021
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer received information alleging an issue related to a continuous positive airway pressure (cpap) device's sound abatement foam.There was no report of patient harm/injury.This issue was reported to the fda per 21 cfr 806.The device will be corrected per res 88058.
 
Additional Manufacturer Narrative
The manufacturer previously reported an allegation of an issue related to sound abatement foam.   a correction to b5 was made and should be:        the manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer previously reported an allegation of an issue related to a cpap device's sound abatement foam and became degraded and the patient has alleged to particles in tubing/chamber.There was no report of patient harm or injury.Repeated attempts to have the device and components returned for evaluation and investigation were unsuccessful.The manufacturer believes they will be unable to gather additional information.The manufacturer is submitting a final report at this time.If pertinent information becomes available to the manufacturer at a later date, an addendum to this final report will be filed.Section h6 updated/corrected in this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REMSTAR SE, W/ HEATED TUBE, SD
Common Device Name
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15206
2673970028
MDR Report Key13115819
Report Number2518422-2021-08890
Device Sequence Number14608551
Product Code BZD
UDI-Device Identifier00606959012509
UDI-Public00606959012509
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K130077
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 05/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberDS220TS
Device Catalogue NumberDS220TS
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/20/2021
Supplement Date Received by Manufacturer05/26/2023
Initial Report FDA Received Date12/29/2021
Supplement Report FDA Received Date05/27/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/17/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
RES 88058
Patient Sequence Number1
Patient SexUnknown
-
-