As the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.Expiry date: 04/2024.
|
|
H10: manufacturing review: a complaint history review was performed.This is the first complaint reported for this product/lot number combination.Therefore, a device history record review is not required.Investigation summary: the sample was not returned for evaluation.The result of the investigation is inconclusive for the reported entrapment of device and material separation issues.The root cause for the reported entrapment of device and material separation issues could not be determined based upon the available information received from the field communications.Labeling review: instructions for use for sleek pta exchange (rx) dilatation catheter product family was reviewed the following sections are applicable: indications: the sleek catheters are intended for balloon dilatation of the femoral, popliteal and infra-popliteal arteries.These catheters are not designed to be used in the coronary arteries.Warnings: visually inspect the packaging to verify that the sterile barrier is intact.Do not use if the sterile barrier is opened or damaged.Do not advance the guidewire, balloon dilation catheter, or any component if resistance is met, without first determining the cause and taking remedial action.This catheter is not recommended for pressure measurement or fluid injection.Precautions: if resistance is felt upon removal, then the balloon, guidewire and the sheath/ guide catheter should be removed together as a unit, particularly if balloon rupture or leakage is known or suspected.This may be accomplished by firmly grasping the balloon catheter and sheath/guide catheter as a unit and withdrawing both together, using a gently twisting motion combined with traction.Before removing catheter from sheath/guide catheter it is very important that the balloon is completely deflated.Proper functioning of the catheter depends on its integrity.Care should be used when handling the catheter.Damage may result from kinking, stretching, or forceful wiping of the catheter.If the hypotube kinks prior to or during use the catheter should be discarded.No attempt should be made to straighten a kink in the hypotube.Directions for use: if any resistance is felt, or if any stretching of the catheter is observed while removing the balloon sleeve, the product should not be used.The catheter should then be inspected for bends, kinks or stretched portions.Do not use if product damage is evident.Deflation and withdrawal: deflate the balloon by drawing a vacuum with a 20 ml or larger syringe.Note: the larger the syringe diameter, the greater the suction that is applied.For maximum deflation a 50 ml syringe is recommended.Simultaneously withdraw the dilatation catheter and guidewire from the guiding catheter/sheath.As the balloon exits the vessel, use a smooth, gentle, steady, motion.If resistance is felt upon removal then the balloon, guidewire and the guiding catheter/sheath should be removed together as a unit under fluoroscopic guidance, particularly if balloon rupture or leakage is known or suspected.This may be accomplished by firmly grasping the balloon catheter and guiding catheter/sheath as a unit and withdrawing both together, using a gentle twisting motion combined with traction.Withdraw the guiding catheter/sheath from the vessel.Apply pressure to the insertion site according to standard practice or hospital protocol for percutaneous vascular procedures.H10: d4 (expiry date: 04/2024), h6 (method).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text: device not returned.
|