Catalog Number 16-02-80
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/02/2021
Event Type  malfunction  
Event Description
Livanova (b)(4) received a report that, the laboratory results show contamination with bacteria outside the limits.The results of the pathological bacteria are below the limits.According to follow up information ifu are respected.There was no patient involvement.
 
Manufacturer Narrative
There was no known patient involvement.The heater-cooler 16-02-80 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).Livanova (b)(4) implemented a field safety notice for disinfection and cleaning of heater-cooler devices.The z number is (b)(4).Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).A complaints database review revealed other previous device contamination complaint submitted by this hospital.According to follow up with the customer,ifu are respected: the device cleaned regularly, no reusable blankets are used, the water quality monitoring is applied and the h2o2 is checked every day, h2o2 is added when the water in the tanks is changed, the disposable water filter is used for tap water, surfaces and water circuits disinfected, prior and after every operation, aerosol collection set is replaced after the allowed use period, device is placed inside the operation theater at nearly 2 meters and with the fan of device is positioned at the opposite of the surgery area.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80309
GM  80309
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w 65th way
arvada, CO 80004
MDR Report Key13116606
MDR Text Key284695223
Report Number9611109-2021-00749
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeFR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial
Report Date 12/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number16-02-80
Was the Report Sent to FDA? No
Date Manufacturer Received12/02/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/01/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-2076/2081-2015
Patient Sequence Number1
-
-