A physician reported a certas valve was implanted via l-p shunt on (b)(6) 2021, with setting 7.
The valve was used with the silascon lumbar catheter (manufactured by kaneka, product code: 702-jj).
On (b)(6) 2021, the setting pressure was changed from 7 to 6.
On an unknown day, the physician tried to change the set pressure because the patient's consciousness was not clear, but the set pressure could not be changed.
On (b)(6), the physician tried to puncture the valve but couldn¿t, and the physician couldn't get the cerebrospinal fluid.
It was suspected shunt dysfunction, therefore, the valve was removed and replaced on (b)(6) 2021.
|