MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 GROSHONG NXT WITH STRIPED CATH 4F SINGLE LUMEN BASIG JAPAN HANG; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 GROSHONG NXT WITH STRIPED CATH 4F SINGLE LUMEN BASIG JAPAN HANG; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number N/A
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/07/2021
Event Type  malfunction  
Manufacturer Narrative
The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of refr2319 showed no other similar product complaint(s) from this lot number.
 
Event Description
It was reported that when the catheter was flushed after connecting the catheter and the oversleeve, leakage was found near the oversleeve.When the new catheter connector was connected to the catheter, no leakage was found.There was no reported patient injury.
 
Event Description
It was reported that when the catheter was flushed after connecting the catheter and the oversleeve, leakage was found near the oversleeve.When the new catheter connector was connected to the catheter, no leakage was found.There was no reported patient injury.
 
Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, complaint and lot history review, applicable previous investigation(s), sample (if available), applicable manufacture records, and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of confirmed is confirmed.One 4 fr single lumen groshong nxt clearvue catheter was returned for evaluation.An initial visual observation showed a small amount of use residue on the returned sample.The returned sample was flushed and pressurized, and a leak was observed in the catheter emanating from within the assembled two-piece connector.The two-piece connector was unassembled, and a microscopic observation revealed a relatively large, diagonal split in the catheter just distal to the metal cannula of the proximal connector piece the catheter was assembled onto.The outer edges of the split were observed to be jagged with a distinct and consistent pattern.The fracture surfaces were observed to be uneven but granular in texture.The shape and texture of the split observed in the returned sample indicate the catheter was most-likely damaged due to contact with a metallic instrument such as hemostats or another hard, textured surface.H3 other text : evaluation findings are in section h.11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GROSHONG NXT WITH STRIPED CATH 4F SINGLE LUMEN BASIG JAPAN HANG
Type of Device
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
johanna de oliveira
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key13117758
MDR Text Key282965806
Report Number3006260740-2021-05479
Device Sequence Number1
Product Code LJS
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K034020
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number7617405J
Device Lot NumberREFR2319
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/10/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/13/2021
Initial Date FDA Received12/30/2021
Supplement Dates Manufacturer Received02/24/2022
Supplement Dates FDA Received02/25/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-