Model Number FGS-0635 |
Device Problem
Positioning Failure (1158)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/09/2021 |
Event Type
malfunction
|
Event Description
|
According to the reporter, they had a capsule which failed to attach to the patient¿s esophagus.They placed another capsule in the same procedure and it attached successfully.There was no patient and user harm.Lubrication was used to facilitate the placement of the capsule.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, they had a capsule which failed to attach to the patient¿s esophagus.They placed another capsule in the same procedure and it attached successfully.There was no patient and user harm.Lubrication was used to facilitate the placement of the capsule.
|
|
Event Description
|
According to the reporter, they had a capsule which failed to attach to the patient¿s esophagus and still attached to the delivery system upon removal.They placed another capsule in the same procedure and it attached successfully.There was no patient and user harm.Lubrication was used to facilitate the placement of the capsule.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Manufacturer Narrative
|
Evaluation summary: medtronic conducted an investigation based upon all information received.The capsule was not returned, but the delivery system was available for evaluation.Visual inspection noted that the plunger was rotated more than one-eighth of a turn.It was reported that the capsule failed to attach to the patient's esophagus.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.The instructions included with this device provide the following guidance: pressing down on the plunger too slowly may result in the capsule not properly attaching to the patient¿s esophagus or not detaching from the delivery device.Do not rotate the plunger while depressing it.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|