• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH DADE INNOVIN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH DADE INNOVIN Back to Search Results
Model Number DADE INNOVIN
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/07/2021
Type of Reportable Event Malfunction
Event or Problem Description
Discordant, falsely elevated prothrombin time (pt) and prothrombin time international normalized ratio (inr) results were obtained on a patient sample on a sysmex ca-660 system using dade innovin reagent.The sample was repeated a second time for pt and inr, recovering lower.The sample was repeated a third time for pt and inr, recovering lower than both of the previous two runs.The sample was then respun and was repeated a fourth time for pt and inr, recovering lower than the three previous runs.The pt and inr results from the third run were reported, as the correct results, to the physician(s).There are no known reports of patient intervention or adverse health consequences due to the falsely elevated prothrombin time (pt) and prothrombin time international normalized ratio (inr) results.
 
Additional Manufacturer Narrative
A united states (us) customer contacted a siemens customer care center (ccc).Quality controls (qc) recovered in range at the time of the event.The issue was limited to one patient sample.The customer suspects that the issue is sample specific.Siemens is investigating the issue.
 
Additional Manufacturer Narrative
Siemens filed the initial mdr 9610806-2021-00079 on 30-dec-2021.Additional information (05-jan-2022): the customer declined to provide further information such as patient printouts, quality control (qc) results, sample handling and centrifugation procedures, and patient medications.The customer stated that they believe the issue was sample specific and not an instrument issue.Inadequate sample collection, mixing, centrifugation or other mishandling of the sample cannot be ruled out as contributing factors to the flagged results.The cause of the event is unknown.The reagent is performing according to specifications.No further evaluation of this device is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DADE INNOVIN
Common Device Name
DADE INNOVIN
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH
emil-von-behring- str. 76
marburg, D-350 41
GM  D-35041
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH
emil-von-behring-str. 76
marburg, D-350 41
GM   D-35041
Manufacturer Contact
christopher aebig
511 benedict ave
tarrytown, NY 10591
9144153450
MDR Report Key13118266
Report Number9610806-2021-00079
Device Sequence Number4998506
Product Code GJS
UDI-Device Identifier00630414023472
UDI-Public00630414023472
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K974343
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/29/2023
Device Model NumberDADE INNOVIN
Device Catalogue Number10873566
Device Lot Number549788
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/07/2021
Supplement Date Received by Manufacturer01/05/2022
Initial Report FDA Received Date12/30/2021
Supplement Report FDA Received Date01/25/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-