MEDTRONIC NEUROMODULATION IMPLANTABLE NEUROSTIMULATOR; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS
|
Back to Search Results |
|
Model Number NEU_INS_STIMULATOR |
Device Problems
Low impedance (2285); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/23/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that the test resistance of the lead during surgery was found to be low.The surgery was completed by replacing the lead the same day and the issue was resolved.
|
|
Manufacturer Narrative
|
Concomitant medical products: product id: 3389s-40, lot# va2tgll9 , product type: lead.Section d information references the main component of the system.Device(s) are: product id: 3389s-40, serial/lot #: (b)(4).If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
D10: section d information references the main component of the system.Other relevant device(s) are: product id 3389s-40 lot# va2tgll9 serial# implanted: explanted: product type lead product id 3389s-40 lot# serial# unknown implanted: explanted: product type lead h3.Analysis of lead (lot# va2tgll9) and a second lead (lot#: unknown) found that there was damage to the outer insulation of the lead body at the depth stop site.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|