DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-0800-53 |
Device Problem
Appropriate Term/Code Not Available (3191)
|
Patient Problems
Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/08/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that during preventive maintenance (pm) performed by a getinge field service engineer (fse) that the cardiosave intra-aortic balloon pump (iabp) displayed ¿internal communication failure¿.The fse additionally observed error codes 124 and 58 in the iabp's error log.There was no patient involvement, and no adverse event was reported.
|
|
Manufacturer Narrative
|
Testing of actual/suspected device (10/3233): the getinge field service engineer (fse) who encountered the issue replaced the front end board.The fse completed the pm and the unit passed all calibration, functional and safety tests.The unit was returned to customer and cleared for clinical use.A supplemental report will be submitted upon completion of our investigation.(b)(6).
|
|