SMITH & NEPHEW, INC. TWINFIX TI 3.5 ULTRABRAID; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
|
Back to Search Results |
|
| Model Number |
72200752 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/10/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).(b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported that during a shoulder cuff repair surgery, the suture of the twinfix titanium 3.5 ultrabraid was breaking up.The procedure was completed with non-significant delay using the same device.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: part of the reported device was received for evaluation.There was no way to determine if the device contributed to the reported event.A visual inspection of the returned device found that it is not in its original packaging.The device has been deployed, and the sutures were not returned.There is debris on the anchor and shaft of the device.Based on the condition of the product material found during visual inspection, additional material testing is not required.The complaint was not confirmed, and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.A complaint history review concluded this was a repeat issue.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the sutures found that a material certification is required with each lot.Factors that could have contributed to the reported event include excessive force or torsion during use or use of sharp instruments near the device tip.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
|
| |
|
Search Alerts/Recalls
|
|
|