| Model Number |
777724 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/15/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the ureteral stent was broken the coil separating from the straight site upon opening the package.It was mentioned that a new ureteral stent had been unpacked for use.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed; however, the cause was unknown.The physical sample was not returned; however, photo samples were submitted by the customer.The break appeared to occur at the large bend on the stent.The break appears to be smooth with no stretching or discoloration of the material, which is similar to what is seen when a stent id cut with a sharp object.A potential root cause for this event could be, "inappropriate package design".The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "for single use only.Do not resterilize.Do not use if the package or product is damaged." "improper handling technique can seriously weaken the stent." "exercise care.Tearing of the stent can be caused by sharp instruments." "care should be exercised when removing the stent from inner polybag so as not to cause tearing or fragmentation." the actual/suspected device was inspected.
|
| |
|
Event or Problem Description
|
|
It was reported that the ureteral stent was broken the coil separating from the straight site upon opening the package.It was mentioned that a new ureteral stent had been unpacked for use.
|
| |
|