• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® X-FORCE® NEPHROSTOMY BALLOON DILATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® X-FORCE® NEPHROSTOMY BALLOON DILATION CATHETER Back to Search Results
Model Number 995101
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/20/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the balloon in the dilation catheter was leaking the fluid and would not fully inflate.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Additional Manufacturer Narrative
The reported event was inconclusive because no sample was returned for evaluation.A potential root cause for this event could be, "improper material selection (stiffness, material impact resistance, aging stability)".The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "if resistance is felt when removing either the catheter or the guidewire from the working sheath, stop and consider removing them as a single unit to prevent damage to the product.Applying excessive force to the catheter can result in tip breakage or balloon separation." "only a physician who has an understanding of the clinical applications, technical principles and associated risks associated with balloon dilation of the nephrostomy tract should use this device." "the x-force® nephrostomy balloon dilation catheter is a sterile, single use device.Carefully inspect the catheter and the sterile packaging for signs of damage that may have occurred during shipment.Do not use the product if damage is evident." "store in a cool, dry and dark place.Do not store near radiation or ultraviolet light sources as these may damage product materials." the device was not returned.
 
Event or Problem Description
It was reported that the balloon in the dilation catheter was leaking the fluid and would not fully inflate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® X-FORCE® NEPHROSTOMY BALLOON DILATION CATHETER
Common Device Name
BALLOON DILATION CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key13120948
Report Number1018233-2021-08589
Device Sequence Number4562517
Product Code LJE
UDI-Device Identifier00801741080722
UDI-Public(01)00801741080722
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K063632
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/20/2023
Device Model Number995101
Device Catalogue Number995101
Device Lot NumberBMEQFM01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/20/2021
Supplement Date Received by Manufacturer04/29/2022
Initial Report FDA Received Date12/30/2021
Supplement Report FDA Received Date05/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/25/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-