• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK AG, BUELACH, SWITZERLAND ORSIRO (US) 2.75/40; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK AG, BUELACH, SWITZERLAND ORSIRO (US) 2.75/40; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number 404674
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/09/2021
Type of Reportable Event Malfunction
Event or Problem Description
An orsiro drug-eluting stent system was selected for treatment of a severely calcified lesion (90 percent stenosis degree) in a moderately tortuous proximal lad and an isr in the distal lad.After pre-dilatation an orsiro stent was successfully implantation in the proximal lad.The second orsiro 2.75/40 was selected to treat the distal isr but could not pass through the already implanted stent and it was decided to remove it for further balloon dilatations.Since the balloon could not pass it was decided that the lesion was too tight and the guidewire was removed.During the control angio it was noticed that the stent has been dislodged from the balloon by getting caught at the guiding catheter.The physician tried to remove the stent with a snare.The patient was sent to cv surgery.The patient was finally doing fine.
 
Additional Manufacturer Narrative
Neither the complaint instrument nor the angiographic material was returned for analysis.Therefore, no technical investigation on the subject could be performed.The product release documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.One photograph taken from the angiogram was provided for analysis.The photograph with a rather poor quality shows a dislodged stent at the distal end of the guiding catheter up to the previously implanted stent whereas the stent appears severely deformed.Review of the product release documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.As a part of final inspection every stent system undergoes visual inspection to ensure correct embedding and homogeneous crimping of the stent.Further, the stent outer diameter and the stent retention force of a defined number of samples is tested from every lot.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause could be determined.The root cause for the complaint event is most likely related to the patients anatomy (i.E.Previously implanted stent).It should be noted that, according to the ifu, when treating multiple lesions, the distal lesion should be initially stented followed by stenting of the proximal lesion.In addition, during slow pull back of the stent system the user is advised to perform fluoroscopic control upon entry of the stent into the guiding catheter to avoid dislodgement of the stent from its position on the delivery system balloon.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORSIRO (US) 2.75/40
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key13122155
Report Number1028232-2021-06989
Device Sequence Number4518216
Product Code NIQ
UDI-Device Identifier07640130439436
UDI-Public07640130439436
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P170030
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/14/2023
Device Model Number404674
Device Catalogue NumberSEE MODEL NO.
Device Lot Number06216272
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/29/2021
Supplement Date Received by Manufacturer01/19/2022
Initial Report FDA Received Date12/30/2021
Supplement Report FDA Received Date01/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-