Catalog Number 326631 |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/03/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that 1 bd insulin syringe with the bd ultra fine¿needle hub separated from the device.The following information was provided by the initial reporter : the customer reported that the needle with the shield was separated from the barrel when the hcp removed the shield.
|
|
Manufacturer Narrative
|
Initial reporter phone# : (b)(6).Initial reporter zip code : (b)(6).A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
|
Event Description
|
It was reported that 1 bd insulin syringe with the bd ultra fine¿needle hub separated from the device.The following information was provided by the initial reporter : the customer reported that the needle with the shield was separated from the barrel when the hcp removed the shield.
|
|
Manufacturer Narrative
|
H6: investigation summary: no samples were returned therefore the investigation was performed based on the photos provided.Photos of a 0.3ml bd insulin syringe were provided.The customer reported that the needle with the shield was separated from the barrel when the hcp removed the shield.The photos were examined, and it was observed that the needle hub/shield assembly was detached from the barrel.No damage to the barrel tip was observed.Bd was able to confirm the customer¿s indicated failure based on the photos received.Capa pr1630423 has been opened to address this issue h3 other text : see h10.
|
|