• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION EMERGE PUSH; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION EMERGE PUSH; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 7138
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Hematoma (1884)
Date of Event 10/26/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported via medwatch mw5105145 that balloon detachment occurred, requiring surgery for removal vascular presented for left heart catheterization via the right radial artery.Disease was noted in the mid left anterior descending artery (lad) and the diagonal branch.The physician ballooned the lad five times with a 2.00mm x 20mm emerge push balloon.The 2.5mm x 15mm emerge balloon was advanced without difficulty, no excess force was noted but while moving into the diagonal, the balloon detached in the artery.An attempt was made to snare the balloon but it was unsuccessful.The physician attempted to inflate the balloon in the guide in order to remove the detached balloon but this was also unsuccessful.Inadvertently the 2.00 x 20mm emerge push balloon detached in the guide in the right arm.The patient became hemodynamically unstable and an intra-aortic balloon pump was inserted in right femoral artery.The patient received two stents to the mid lad and a ventricular assist device was inserted and then removed.The patient was transported to a higher level of care for urgent retrieval of the balloons.The patient underwent vascular surgery and both balloons were retrieved.The distal diagonal hematoma was also treated as well as right artery ligation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMERGE PUSH
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13122409
Report Number2134265-2021-16372
Device Sequence Number19988750
Product Code LOX
UDI-Device Identifier08714729806011
UDI-Public08714729806011
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K121196
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/29/2023
Device Model Number7138
Device Catalogue Number7138
Device Lot Number0027241842
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/20/2021
Initial Report FDA Received Date12/30/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/03/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-