• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SPINE INC SYMPHONY OCT SYSTEM ROD CUTTER FOR 3.5MM AND 4.0MM RODS; INSTRUMENT, CUTTING, ORTHOPEDIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SPINE INC SYMPHONY OCT SYSTEM ROD CUTTER FOR 3.5MM AND 4.0MM RODS; INSTRUMENT, CUTTING, ORTHOPEDIC Back to Search Results
Model Number 2020-00-135
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/03/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that on (b)(6) 2021, during an unknown procedure a crack in the tip of the rod cutter was noticed.The issue was found after it was used to cut a rod.There were no fragments generated.Procedure was successfully completed with no surgical delay.There was no patient consequences.There is no further information available.This report is for one (1) symphony oct system rod cutter for 3.5mm and 4.0mm rods.This is report 1 of 1 for (b)(4).
 
Additional Manufacturer Narrative
The device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H3, h6: a product investigation was conducted.Visual inspection: the complaint device rod cutter 3.5 and 4.0 rods (product code: 202000135, lot number: gs676412) was returned to cq west chester for investigation.The rod cutter had a crack starting from the jaw to the opening for the 4.0 mm rods.Also, the etch on the item was fading.The item was discolored.Document /specification review: based on the date of manufacture, the current and manufactured version of the drawings were reviewed.Dimensional inspection: complaint relevant dimension could not be measured due to the design of the device.Complaint can be confirmed based on the available information.Conclusion: the rod cutter had a crack on its jaw and was discolored.A definitive root cause was not determined during investigation.There was no indication that a design or manufacturing issue contributed to the complaint.No design issues were observed during the document/specification review.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.H3, h6: a device history record (dhr) review was conducted: a review of the receiving inspection (ri) for rod cutter 3.5 and 4.0 rods was conducted identifying that lot number gs676412 was released in two (2) batches.Batch1: lot qty of 212 units were released on oct 28, 2019 with no discrepancies.Batch2: lot qty of 1 units were released on dec 5, 2019 with no discrepancies.As a result, the ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYMPHONY OCT SYSTEM ROD CUTTER FOR 3.5MM AND 4.0MM RODS
Common Device Name
INSTRUMENT, CUTTING, ORTHOPEDIC
Manufacturer (Section D)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
BRIDGEWATER DISTRIBUTION
50 scotland boulevard
bridgewater MA 02324
Manufacturer Contact
kara ditty-bovard
chemin-blanc 38
le locle 02400
SZ   02400
6103142063
MDR Report Key13129386
Report Number1526439-2021-02681
Device Sequence Number1844231
Product Code HTZ
UDI-Device Identifier10705034530536
UDI-Public10705034530536
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2020-00-135
Device Catalogue Number202000135
Device Lot NumberGS676412
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/20/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/06/2021
Supplement Date Received by Manufacturer01/28/2022
Initial Report FDA Received Date12/31/2021
Supplement Report FDA Received Date02/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/05/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-