• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MENTOR TEXAS MENTOR SILTEX CONTOUR PROFILE SPECTRUM; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR SILTEX CONTOUR PROFILE SPECTRUM; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE Back to Search Results
Model Number 3542513
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Deformity/ Disfigurement (2360)
Date of Event 11/26/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: left deflation.Explantation date: (b)(6) 2022.(b)(4).
 
Event or Problem Description
It was reported that a (b)(6)-year-old caucasian female patient underwent revision breast reconstruction surgery with a 450cc mmentor siltex contour profile spectrum and experienced breast implant deflation on her left side postoperatively.As a result, the device will be explanted on (b)(6) 2022.
 
Additional Manufacturer Narrative
On march 22, 2022, device evaluation was completed.Device evaluation summary: the product was returned to mentor for evaluation.The tubing, the connector, and the injection port were not received.Mentor conducted a visual inspection, leak testing, and microscopic examination of the returned device.Visual analysis of the returned sample revealed that no damage or anomalies were observed on the spec siltex contour 450cc breast implant.Leak testing was performed, in accordance with mentor procedures, and a tear was noted on the posterior view measuring approximately 0.7 cm.A microscopic examination was performed and the cause of the deflation could not be identified.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Although no conclusion could be reached on the cause of the reported event, the instructions for use contain the following caution: causes of deflation of saline implants include but are not limited to the following events: crease fold failure, intraoperative or postoperative trauma, excessive stresses or manipulations as may occur during normal, daily routines including customary and purposeful trauma such as that which can occur during vigorous exercise, athletics, etc.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On (b)(6) 2022, the mentor failure analysis lab received the device for evaluation.The analysis has begun, but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR SILTEX CONTOUR PROFILE SPECTRUM
Common Device Name
PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
gabriel alfageme
3041 skyway circle north
irving, TX 75038
9497898687
MDR Report Key13137748
Report Number1645337-2022-00006
Device Sequence Number1867372
Product Code FWM
UDI-Device Identifier00081317002154
UDI-Public00081317002154
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P990075
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/09/2012
Device Model Number3542513
Device Catalogue Number3542513
Device Lot Number5861101
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/21/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/08/2021
Supplement Date Received by Manufacturer02/21/2022
03/22/2022
Initial Report FDA Received Date01/01/2022
Supplement Report FDA Received Date03/07/2022
03/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/10/2008
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age62 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-