MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ENDOTAK RELIANCE SG; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ENDOTAK RELIANCE SG; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT) Back to Search Results
Model Number 0171
Medical Device Problem Code High impedance (1291)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/12/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This was not returned.As such, physical analysis has not been conducted in our laboratory.However, a labeling review and risk review were conducted.These reviews determined that the clinical observation of high impedance is a known inherent risk of the lead.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that this right ventricle (rv) lead exhibited high, out of range shock impedance of great than 200 ohms.The physician reported the shock impedances to be consistently high.All other measurements are within normal range.The physician will monitor the patient closely.The device remains implanted.No adverse patient effects were reported.New information was received during a follow up appointment, lead integrity testing , at 1,1 j and shock impedance was >125ohm.The physician scheduled the patient for a lead revision procedure the next day.New information was received that this rv lead was explanted, and new one implanted.X-ray images confirmed no lead fractures.Besides surgical intervention, no adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOTAK RELIANCE SG
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key13138627
Report Number2124215-2021-38930
Device Sequence Number13910969
Product Code LWS
UDI-Device Identifier00802526373374
UDI-Public00802526373374
Combination Product (Y/N)Y
Initial Reporter CountryIT
PMA/510(K) Number
P910073/S041
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2005
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date01/21/2007
Device Model Number0171
Device Catalogue Number0171
Device Lot Number100580
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/17/2021
Initial Report FDA Received Date01/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/21/2005
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
-
-