• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SOFRADIM PRODUCTION SAS SYMBOTEX; MESH, SURGICAL, POLYMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SOFRADIM PRODUCTION SAS SYMBOTEX; MESH, SURGICAL, POLYMERIC Back to Search Results
Model Number SYM2515E
Medical Device Problem Codes Migration or Expulsion of Device (1395); Mechanics Altered (2984); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Abdominal Pain (1685); Adhesion(s) (1695); Emotional Changes (1831); Fever (1858); Hematoma (1884); Inflammation (1932); Nausea (1970); Nerve Damage (1979); Pain (1994); Pneumonia (2011); Seroma (2069); Vomiting (2144); Hernia (2240); Anxiety (2328); Distress (2329); Numbness (2415); Obstruction/Occlusion (2422); Respiratory Failure (2484); Ascites (2596); Abdominal Distention (2601); Fluid Discharge (2686); Unspecified Tissue Injury (4559); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Type of Reportable Event Serious Injury
Event or Problem Description
The patient¿s attorney alleged a deficiency against the device.The product was used for therapeutic treatment of a ventral incisional hernia.It was reported that after the implant, the patient experienced mental pain/anguish, pain, disability, impairment, loss of enjoyment of life, defective mesh, and emotional distress.Post-operative patient treatment included medical treatment.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Additional info: b2 (added hospitalization and life threatening), ( ime e2402: "emp hysema, abnormal wbc, potassium, and chloride, ileus, leukocytosis, cord lipoma").Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The patient¿s attorney alleged a deficiency against the device.The product was used for therapeutic treatment of a ventral incisional hernia.It was reported that after the implant, the patient experienced bulge, abdominal pain, serosanguinous fluid discharge, cord lipoma, pneumonia, acute respiratory failure, fever, nausea, vomiting, emphysema, abnormal wbc, potassium and chloride, ileus, leukocytosis, adhesions, obstruction, recurrence, nerve damage, seroma, inflammation, hematoma, fluid collection, anxiety, numbness, mesh migration, suffering, balled up mesh, mental pain/anguish, pain, disability, impairment, loss of enjoyment of life, defective mesh, and emotional distress.Post-operative patient treatment included ultrasound, procedure for right upper quadrant fluid collection, ultrasound guided right anterior abdominal wall fluid collection aspiration, lap lysis of adhesions, bilateral component separation, transverse abdominis release, ct scan, abdominal x-rays, gastrografin, hospitalization, ng tube placement, iv fluids, iv pain medications, iv antibiotics, additional surgery for hernia repair, closure of posterior rectus sheath defect and placement of overlying retrorectus tissue reinforcement, total parenteral nutrition, physical therapy, medical treatment and hernia repair with new mesh.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The patient¿s attorney alleged a deficiency against the device.The product was used for therapeutic treatment of a ventral incisional hernia.It was reported that after the implant, the patient experienced adhesions, obstruction, recurrence, nerve damage, seroma, inflammation, hematoma, fluid collection, anxiety, numbness, mesh migration, suffering, balled up mesh, mental pain/anguish, pain, disability, impairment, loss of enjoyment of life, defective mesh, and emotional distress.Post-operative patient treatment included medical treatment and hernia repair with new mesh.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYMBOTEX
Common Device Name
MESH, SURGICAL, POLYMERIC
Manufacturer (Section D)
SOFRADIM PRODUCTION SAS
116 avenue du formans
trevoux 01600
FR  01600
Manufacturer (Section G)
SOFRADIM PRODUCTION SAS
116 avenue du formans
trevoux 01600
FR   01600
Manufacturer Contact
justin ellis
5920 longbow drive
8200 coral sea st ne
mounds view,mn, CO 55112
7635265677
MDR Report Key13139274
Report Number9615742-2022-00002
Device Sequence Number1686870
Product Code FTL
UDI-Device Identifier10884521190474
UDI-Public10884521190474
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K142908
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2022
Device Model NumberSYM2515E
Device Catalogue NumberSYM2515E
Device Lot NumberPRI0177X
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/15/2021
Supplement Date Received by Manufacturer02/08/2023
02/22/2023
Initial Report FDA Received Date01/02/2022
Supplement Report FDA Received Date02/13/2023
03/06/2023
Date Device Manufactured10/26/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Life Threatening; Disability; Required Intervention;
Patient SexMale
Patient Weight136 KG
Patient EthnicityHispanic
-
-