| Model Number |
120707S |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Cardiac Tamponade (2226)
|
| Date of Event |
12/08/2021
|
|
Type of Reportable Event
|
Death
|
|
Additional Manufacturer Narrative
|
|
Device investigation details: information received indicates that the device is not available for return, therefore no product investigation can be performed, and the customer complaint cannot be confirmed.Investigation is in progress, once completed a supplemental will be submitted.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.(b)(6).If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that an (b)(6) male patient ((b)(6)) underwent a clinical trial to inject a study drug into a muscle in the left ventricle with a noga-star¿ cardiology catheter.The patient suffered cardiac tamponade requiring pericardiocentesis and surgical intervention and subsequently death.The patient died after the procedure.The patient had already moved to the ward hours earlier.No abnormal or unusual symptoms were observed during the procedure.Mapping was performed with a noga-star¿ cardiology catheter and drug injections with a myostar catheter.That patient was part of a multinational gene therapy study.Cardiac tamponade was observed in the patient on ultrasound.An ultrasound examination was performed as the patient's condition worsened.The adverse event was discovered post use of biosense webster products about 1.5 hours after the procedure.Medical intervention was provided after first symptoms, transthoracic echocardiography, pericardiocentesis, resuscitation, open heart surgery.Physician¿s opinion on the cause of this adverse event was reported as procedure related, not bwi product malfunction and also patient condition related since myocardium was abnormal; in the surgeon's opinion.Relevant patient history included myocardial infarction.No generator was in use or connected as no ablation were performed.The event occurred post procedure.The date of death was(b)(6) 2021.The physician¿s opinion, what was the cause of death: not yet confirmed.Forensic autopsy will confirm the cause of death.In the open-heart surgery, the surgeon found 4-5 small holes in the lateral wall of the left ventricle.Drugs had been injected into the same area.The surgeon tried to close them with sutures without success.Suturing was not successful because the tissue could not withstand suturing.The left ventricle of the heart is mapped with a nogastar catheter.Based on the pet scan, the area where the study drug is to be injected is decided.Injections are performed with a myostar catheter.The study medicine contains stem cells.They are intended to grow new blood vessels in that area and improve heart function.The myostar catheter needle is adjusted to 5mm out of the catheter tip.There was no post-op device malfunction reported.
|
| |
|
Additional Manufacturer Narrative
|
|
On 3-jan-2022, additional information was received which confirmed the cause of death was the pericardium effusion that was caused by the procedure because the tissue was found to be abnormal and suturing was impossible.This may be due to scarring of the heart muscle and the muscle was attached to the pericardium.Muscle damage increased when the surgeon opened the pericardium.The autopsy assumed it that way.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
It was reported that an 83-year-old male patient (80kg) underwent a clinical trial to inject a study drug into a muscle in the left ventricle with a noga-star¿ cardiology catheter.The patient suffered cardiac tamponade requiring pericardiocentesis and surgical intervention and subsequently death.The patient died after the procedure.The patient had already moved to the ward hours earlier.No abnormal or unusual symptoms were observed during the procedure.Mapping was performed with a noga-star¿ cardiology catheter and drug injections with a myostar catheter.That patient was part of a multinational gene therapy study.Cardiac tamponade was observed in the patient on ultrasound.An ultrasound examination was performed as the patient's condition worsened.The adverse event was discovered post use of biosense webster products about 1.5 hours after the procedure.Medical intervention was provided after first symptoms, transthoracic echocardiography, pericardiocentesis, resuscitation, open heart surgery.Physician¿s opinion on the cause of this adverse event was reported as procedure related, not bwi product malfunction and also patient condition related since myocardium was abnormal; in the surgeon's opinion.On 3-jan-2022, additional information was received which confirmed the cause of death was the pericardium effusion that was caused by the procedure because the tissue was found to be abnormal and suturing was impossible.This may be due to scarring of the heart muscle and the muscle was attached to the pericardium.Muscle damage increased when the surgeon opened the pericardium.The autopsy assumed it that way.Device investigation details: the device investigation has been completed as it has been determined that the lot number provided by the customer (lot # 3036774m) was not recognized as a valid lot number in our system.As such, a manufacturing record evaluation (mre) cannot be conducted.Still no device has been received for analysis, therefore, no product investigation can be performed, and the customer complaint cannot be confirmed.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer's ref.# (b)(4) corrections: on 31-mar-2022, it was noticed that incorrect dates were reported in previous supplemental medwatch reports.Supplemental (follow-up # 1): in field g3.Date received by manufacturer was reported as 3-jan-2021 but it should have been 3-jan-2022.Supplemental (follow-up # 2): in field g3.Date received by manufacturer was reported as 26-jan-2021 but it should have been 26-jan-2022.
|
| |
|
Additional Manufacturer Narrative
|
|
On 14-mar-2022, additional information was received indicating it was realized the wrong lot number was reported upon initial report of this event.The correct lot number is 30367744m.Field d4.Lot, has been updated with this information.Per the updated lot number, the manufactured date and expiration date have been provided.Therefore, fields d4.Expiration date and h4.Device manufacture date have been populated as well (manufacturing date: 2020-03-24 ; expiration date: 2023-03-23).A manufacturing record evaluation was performed for the finished device 30367744m number, and no internal actions related to the complaint was found during the review.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.#(b)(4).
|
| |
|
Search Alerts/Recalls
|