It was reported that, during an internal fixation surgery, a t7 driver shaft with ao quick connect snapped while inside the patient.All pieces were recovered.Surgery was resumed, without any delay, with a back-up device.Patient was not injured as consequence of this problem.
|
|
H3, h6: the associated device was not returned for evaluation and the reported event could not be confirmed, therefore, the contribution of the device to the reported event could not be corroborated.The clinical/medical investigation concluded that, per the complaint details, all the pieces were recovered from inside of the patient.Reportedly, the surgery was resumed without any delay with a back-up device.Since there was no alleged harm to the patient, no further clinical/medical assessment is warranted at this time.Should any additional relevant patient information be provided, this case would be re-assessed.A review of complaint history did not reveal similar events for the listed device.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.An historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|