SMITH & NEPHEW, INC. LONG HEXDRIVER; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 71631070
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/14/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, during an internal fixation surgery, while inserting the screw using a long hexdriver, the tip of the inner rod broke off.The piece was remained inside the head of a trigen low profile screw 5.0mm x 30mm (concomitant), and then, was removed along with the screw.Pieces of the broken device fell inside the patient, all were successfully retrieved.Surgery was resumed, after a non-significant delay, with a back-up device and back-up screw.Patient was not injured as consequence of this problem.
 
Additional Manufacturer Narrative
Internal complaint reference (b)(4).
 
Additional Manufacturer Narrative
H3, h6: the associated device was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The device was fractured along the tip of the inner rod, rendering the device inoperative.The device shows signs of extensive use.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.The clinical/medical evaluation concluded that based on the information provided, the tip of the inner rod broke off and was retained in the head of the trigen screw and they were removed in addition to, the screw.Subsequently all the pieces of the broken device that fell inside the patient were all removed.According to the report, the surgery was resumed, after a non-significant delay, with a back-up device and back-up screw.Since there was no other harm alleged to this patient, no further clinical/medical assessment is warranted at this time.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.Due to the need for a backup device, a contribution of the device to the reported event could be corroborated.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.Additional information: d3 and d8/d9.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LONG HEXDRIVER
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13142335
Report Number1020279-2022-00009
Device Sequence Number16520362
Product Code HSB
UDI-Device Identifier03596010519269
UDI-Public03596010519269
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K122170
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71631070
Device Catalogue Number71631070
Device Lot Number15FM17295
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/28/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/14/2021
Supplement Date Received by Manufacturer02/10/2022
Initial Report FDA Received Date01/03/2022
Supplement Report FDA Received Date02/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/23/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
71645030