• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SCREW BIOSURE REGENESORB 8MM X 25MM; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SCREW BIOSURE REGENESORB 8MM X 25MM; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE Back to Search Results
Model Number 72204399
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/14/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that during the procedure, when inserting the screw in the tunnel, the screw got broken.The procedure was completed with a s+n back-up device.No delay was reported, and no other complications were reported.
 
Additional Manufacturer Narrative
H3, h6: the reported device was received for evaluation.There was a relationship found between the device and the reported event.A complaint history review found similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the polymer found that the storage requirements for the material are specified, and the material must comply with provided percentage composition specifications as measured by ash test.A visual inspection of the returned device found that it is not in its original packaging.The screw is fractured on the distal end, and there is debris on the device.Based on the condition of the product material found during visual inspection, additional material testing is not required.No clinically relevant supporting documentation was provided for review.If a portion of the biosure screw remains retained in the patient, it is made of a biocomposite material which is intended for implantation to allow for bone ingrowth.If the screw was retained it is in the tunnel, therefore, the potential risk for tissue inflammation and/or pain, micro-motion/migration is unlikely.Therefore, no further clinical/medical assessment is warranted at this time.The complaint was confirmed, and the root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include excessive force on the device, excessive torque on the device, attempted correction of a damaged device, off-axis insertion, improper preparation of the insertion site, or an inadvertent impact event inconsistent with normal use.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCREW BIOSURE REGENESORB 8MM X 25MM
Common Device Name
FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13142469
Report Number1219602-2022-00010
Device Sequence Number17310065
Product Code MAI
UDI-Device Identifier00885554036756
UDI-Public00885554036756
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K142948
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/28/2024
Device Model Number72204399
Device Catalogue Number72204399
Device Lot Number50918751
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/15/2021
Supplement Date Received by Manufacturer12/27/2022
Initial Report FDA Received Date01/03/2022
Supplement Report FDA Received Date12/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/28/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-