SMITHS MEDICAL ASD, INC. BIVONA; TRACHEOSTOMY TUBE AND TUBE CUFF Back to Search Results
Model Number 67NFP40
Medical Device Problem Code Inflation Problem (1310)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/04/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the cuffs did not inflate when checking prior to the patient trach change.
 
Additional Manufacturer Narrative
Other, other text: h10: device evaluation: samples received: two (2) samples were received from p/n 67nfp40 in used conditions, without original packaging and with its certificate of safe handling.Functional test: samples were filled with 5cc of air according to mp bivona tt-tj130 rev.102 ? "leak test" to see if there was any functional problem.Results: when the sample was inflated with air, the cuff only inflated one side.According to the ifu (110018853-001 rev.100 - instruction for use ? pedi/neo tts ifu) the pilot balloon was manipulated, and the cuff inflated.After the whole cuff inflated, it was deflated and inflated 4 times, all the time the cuff inflated completely and symmetrically, based on analysis the complaint is not confirmed.Root cause could not be determined since the sample successfully passed functional test.No corrective actions were taken since the complaint was not confirmed.A dhr review was performed subsequent to the manufacturing of the device and prior to its release.No problems or issues were identified during this dhr review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIVONA
Common Device Name
TRACHEOSTOMY TUBE AND TUBE CUFF
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer (Section G)
NULL
5700 west 23rd ave
gary IN 46406
Manufacturer Contact
david halverson
6000 nathan lane north
room l-041
minneapolis, MN 55442
MDR Report Key13142897
Report Number3012307300-2022-00015
Device Sequence Number16527344
Product Code JOH
UDI-Device Identifier10351688518767
UDI-Public10351688518767
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K923878
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number67NFP40
Device Lot Number4122439
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/29/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/06/2021
Supplement Date Received by Manufacturer05/03/2022
Initial Report FDA Received Date01/03/2022
Supplement Report FDA Received Date06/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-