It was reported that the femoral peg drill continues to bind in the posterior lug hole of the 2-in-1 cut blocks.Significant metal debris caused by the binding.Lug drill ultimately locked itself in the lug hole, needing significant force to disassociate, causing cut block to shift before making saw cuts.It is our understanding there now exist a new lug drill to avoid this issue.No case involved.
|
|
H3, h6: the associated device was not returned for evaluation and the reported event could not be confirmed.A review of complaint history revealed similar events for the listed device, but no similar events for the batch, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.An historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.The contribution of the device to the reported event could not be corroborated.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|