Model Number 101/860/080CZ
Device Problem Deflation Problem (1149)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/24/2021
Event Type  malfunction  
Event Description
It was reported that the customer started using the product for regular replacement and he noticed the cuff was spontaneously deflated several times with the lapse of time.The customer changed the product to another new one and the device performed normally.No patient injury was reported.
 
Manufacturer Narrative
Other text: one unit was returned for investigation.Upon physical inspection, it was found that the complained issue could be duplicated.The problem source is design.Dhr review was done, no issues related to the original complaint were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX
Type of Device
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
akasaka, minato-ku, tokyo
minneapolis, MN 55442
MDR Report Key13146382
MDR Text Key283685029
Report Number3012307300-2022-00022
Device Sequence Number1
Product Code BTO
UDI-Device Identifier15019517076158
UDI-Public15019517076158
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K173384
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number101/860/080CZ
Device Catalogue Number101/860/080CZ
Device Lot Number4061061
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/14/2021
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/10/2021
Initial Date FDA Received01/03/2022
Supplement Dates Manufacturer Received05/13/2022
Supplement Dates FDA Received06/24/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/30/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-