• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRITANIUM BPLATE TRIATHLON S6; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRITANIUM BPLATE TRIATHLON S6; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Model Number 5536-B-600
Medical Device Problem Codes Fracture (1260); Unstable (1667)
Health Effect - Clinical Code Joint Laxity (4526)
Date of Event 12/09/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Device not returned to the manufacturer.
 
Event or Problem Description
It was reported that the patient's right knee was revised due to malposition and instability.Intra-operatively the following was noted: slight malposition of femoral component, worn insert with broken locking wire, and broken tibial baseplate.The baseplate was broken medially 1/3 of the way through the anterior medial spike.The patient was revised to a ts knee with a universal baseplate and stems.
 
Additional Manufacturer Narrative
An event regarding crack/fracture involving a triathlon baseplate was reported.The event was confirmed via evaluation of the provided photographs of the device.Method & results: -product evaluation and results: the reported device was not returned however photographs were provided for review.Visual inspection of the provided intraoperative and explantation photographs indicated that the baseplate is fractured.The device is covered in blood and tissue.No further observations can be made from the photographs.-clinician review: no medical records were received for review with a clinical consultant.-product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to instability.Fracture of the triathlon baseplate was observed intraoperatively.The reported device was not returned however photographs were provided for review.Visual inspection of the provided intraoperative and explantation photographs indicated that the baseplate is fractured.The device is covered in blood and tissue.No further observations can be made from the photographs.The exact cause of the event could not be determined because insufficient information was provided.Further information such as pathology reports, pre- and post-operative x-rays and the primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
It was reported that the patient's right knee was revised due to malposition and instability.Intra-operatively the following was noted: slight malposition of femoral component, worn insert with broken locking wire, and broken tibial baseplate.The baseplate was broken medially 1/3 of the way through the anterior medial spike.The patient was revised to a ts knee with a universal baseplate and stems.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRITANIUM BPLATE TRIATHLON S6
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-LIMERICK
raheen business park
limerick NA
EI   NA
Manufacturer Contact
joann ripoli
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key13150847
Report Number0002249697-2022-00007
Device Sequence Number14805648
Product Code MBH
UDI-Device Identifier07613327041514
UDI-Public07613327041514
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K123486
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/11/2023
Device Model Number5536-B-600
Device Catalogue Number5536B600
Device Lot NumberCTD21820
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/09/2021
Supplement Date Received by Manufacturer11/04/2022
Initial Report FDA Received Date01/04/2022
Supplement Report FDA Received Date11/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/13/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age71 YR
Patient SexMale
Patient Weight158 KG
-
-