MEDTRONIC IRELAND VALIANT NAVION; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR Back to Search Results
Model Number VNMF3131C97TU
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code Aneurysm (1708)
Date of Event 05/28/2020
Type of Reportable Event Malfunction
Event or Problem Description
A valiant navion was implanted during an endovascular procedure for the treatment of a thoracic aortic aneurysm with a penetrating thoracic ulcer.It was reported the patient returned for a follow-up ct approximately 19 months later where it was discovered that there was aneurysmal dilation/penetrating ulceration near the level of the descending aortic graft which has significantly increased measuring up to 56mm at the proximal margin of the graft with associated endoleak around the graft.The aorta measures up to 51mm at the level of distal margin of the graft with associated endoleak around the distal stent.The aneurysm may have an inflammatory component given some erosion along the anterior margin of the t10 vertebral body.The aorta at the level of the hiatus measures up to 53mm , previously 34mm.No cause of event as per the physician has been provided.No additional clinical sequalae has been provided and the patient is fine.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Corelab's review of ct imaging from approximately 3.5 months, 6 months and 1.5 year post index procedure have identified a type ia endoleak.No fracture, stent ring enlargement or aortic enlargement were identified.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALIANT NAVION
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key13151047
Report Number9612164-2022-00023
Device Sequence Number5479301
Product Code MIH
UDI-Device Identifier00763000100599
UDI-Public00763000100599
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
P100040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/02/2020
Device Model NumberVNMF3131C97TU
Device Catalogue NumberVNMF3131C97TU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/14/2021
Supplement Date Received by Manufacturer01/15/2022
Initial Report FDA Received Date01/04/2022
Supplement Report FDA Received Date02/11/2022
Date Device Manufactured09/03/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age67 YR
Patient SexMale
-
-