ZIMMER SWITZERLAND MANUFACTURING GMBH CMN FEMORAL NAIL, CCD 125, LEFT, 10 MM, 42 CM; ZIMMER NATURAL NAIL SYSTEM CEPHALOMEDULLARY NAILS Back to Search Results
Model Number N/A
Medical Device Problem Codes Fracture (1260); Material Integrity Problem (2978); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 11/24/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that when the nail was tried to get removed it broke at the level of the wings of the cephalic screw, impossible to remove the material that remained in the patient's leg.
 
Additional Manufacturer Narrative
The manufacturer did not receive x-rays, or other source documents for review.The manufacturer did not receive the device for investigation.The lot number of the device was received.The device history records will be reviewed during investigation.A cause for this specific event cannot be ascertained from the information provided.As soon as supplemental information becomes available an updated report will be submitted.Zimmer biomet¿s reference number of this file is (b)(4).
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.Investigation and conclusion 1.Event description: it was reported that: when the nail was removed, it broke at the level of the extensions of the cephalic screw, making it impossible to remove the material that remained in the patient's leg.The znn nail was broken during the removal.Harm: s3 - instability, moderate hazardous situation: user cannot remove existing and well-fixed implants during a medical or surgical intervention.2.Review of received data: no medical data (like surgical reports or x-rays) relevant to the case has been received.Due diligence: further due diligence to support the event description was completed and documented.3.Product evaluation: no product was returned; therefore, visual and dimensional evaluation could not be performed.4.Review of product documentation: - device purpose: this device is intended for treatment.The compatibility check could not be performed as only the complained product is known.Dhr review: review of the device history records identified no deviations or anomalies during manufacturing.Complaint history review: review of complaint history identified no additional similar complaints for the reported item and no additional complaints for the reported part and lot combination related to the event.5.Conclusion: it was reported that: when the nail was removed, it broke at the level of the extensions of the cephalic screw, making it impossible to remove the material that remained in the patient's leg.The znn nail was broken during the removal.The device manufacturing history records indicate that the znn nail was accepted into final stock with no reported discrepancies.Neither x-rays, operative notes, office visit notes, nor device(s) or photos of the device(s) were received; therefore the condition of the component(s) is unknown.Patient factors that may have affected the performance of the components such as bone quality, activity level, type of activity (low impact vs.High impact), and relevant medical history are unknown.Adherence to rehabilitation protocol is unknown.Based on the investigation the reported event cannot be confirmed.The investigation did not identify a nonconformance or a complaint out of box (coob).Based on the given information and the results of the investigation, we were not able to identify a specific root cause for this issue.No corrective actions, preventive actions, or field actions resulted after investigation of this event.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
Investigation results are now available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CMN FEMORAL NAIL, CCD 125, LEFT, 10 MM, 42 CM
Common Device Name
ZIMMER NATURAL NAIL SYSTEM CEPHALOMEDULLARY NAILS
Manufacturer (Section D)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ  8404
Manufacturer (Section G)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ   8404
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key13151591
Report Number0009613350-2022-00005
Device Sequence Number19870369
Product Code HSB
UDI-Device Identifier00889024300187
UDI-Public00889024300187
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K091566
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2025
Device Model NumberN/A
Device Catalogue Number47-2493-421-10
Device Lot Number2794176
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/08/2021
Supplement Date Received by Manufacturer03/15/2022
Initial Report FDA Received Date01/04/2022
Supplement Report FDA Received Date03/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/06/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age65 YR
Patient SexMale
Patient Weight75 KG
-
-