|
This follow-up report is being submitted to relay additional and/or corrected information.Investigation and conclusion 1.Event description: it was reported that: when the nail was removed, it broke at the level of the extensions of the cephalic screw, making it impossible to remove the material that remained in the patient's leg.The znn nail was broken during the removal.Harm: s3 - instability, moderate hazardous situation: user cannot remove existing and well-fixed implants during a medical or surgical intervention.2.Review of received data: no medical data (like surgical reports or x-rays) relevant to the case has been received.Due diligence: further due diligence to support the event description was completed and documented.3.Product evaluation: no product was returned; therefore, visual and dimensional evaluation could not be performed.4.Review of product documentation: - device purpose: this device is intended for treatment.The compatibility check could not be performed as only the complained product is known.Dhr review: review of the device history records identified no deviations or anomalies during manufacturing.Complaint history review: review of complaint history identified no additional similar complaints for the reported item and no additional complaints for the reported part and lot combination related to the event.5.Conclusion: it was reported that: when the nail was removed, it broke at the level of the extensions of the cephalic screw, making it impossible to remove the material that remained in the patient's leg.The znn nail was broken during the removal.The device manufacturing history records indicate that the znn nail was accepted into final stock with no reported discrepancies.Neither x-rays, operative notes, office visit notes, nor device(s) or photos of the device(s) were received; therefore the condition of the component(s) is unknown.Patient factors that may have affected the performance of the components such as bone quality, activity level, type of activity (low impact vs.High impact), and relevant medical history are unknown.Adherence to rehabilitation protocol is unknown.Based on the investigation the reported event cannot be confirmed.The investigation did not identify a nonconformance or a complaint out of box (coob).Based on the given information and the results of the investigation, we were not able to identify a specific root cause for this issue.No corrective actions, preventive actions, or field actions resulted after investigation of this event.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|