| Model Number |
SN6AT5 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Blurred Vision (2137)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.(b)(4).
|
| |
|
Event or Problem Description
|
|
A physician reported that following an intraocular lens (iol) implant procedure, the patient experienced blurred vision.The iol was exchanged for another lens model 2 months following the initial implant procedure.Additional information has been requested.
|
| |
|
Event or Problem Description
|
|
Additional information received and reported that a 5.175 lens was exchanged for a 6.190 lens due to blurry vision and refractive surprise.This does not meet criteria for reporting based on applicable medical device regulations.A lens exchange from one power to a different power is an attempt to adjust or fine tune the refractive outcome trying to achieve a desired target when the initial lens calculations did not achieve the intended outcome.There is no reported defect or fault with the lens.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on information received following submission of the initial report, this event does not meet criteria for reporting as a serious injury.The manufacturer internal reference number is: (b)(4).
|
| |
|