• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES NAVIO LONG ATTACHMENT; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES NAVIO LONG ATTACHMENT; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number PFSR110006
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/15/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference case- (b)(4).
 
Event or Problem Description
It was reported that, during the milling stage in a navio assisted surgery, it was found that there were black fumes coming from the drill.The navio long attachment and exposure control guard became hot.The surgery was continued, without significant delays, using the navio long attachment from other set and with the same technique.The patient was not harmed beyond the reported problem.
 
Additional Manufacturer Narrative
H3, h6: the navio long attachment, p/n pfsr110006, lot 86885, used for treatment was returned for evaluation.A relationship between the reported event and the device was established.Nothing was identified visually that contributed to the reported problem.A functional evaluation was performed.The reported problem was confirmed.Some smoke appeared during testing and some debris can be observed inside the attachment.The most likely cause of this event is due to degradation/disintegration of the long attachment internal bearings.A review of manufacturing records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints found similar events.A historical escalation event review was completed.The review determined that prior escalation actions are applicable to the scope of this complaint however no further escalation action is required.The failure mode and associated risk have been anticipated within the risk file and the documented risk level is still adequate.No changes as a result of investigation.Bearing disintegration is associated with wear and tear over time.Although no further containment or corrective action is recommended or required at this time, the failure mode will continue to be closely monitored through complaint investigation and trended through post market surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVIO LONG ATTACHMENT
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13152351
Report Number3010266064-2022-00007
Device Sequence Number20186558
Product Code OLO
UDI-Device Identifier00885556628607
UDI-Public00885556628607
Combination Product (Y/N)N
Initial Reporter CountryIN
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPFSR110006
Device Catalogue NumberPFSR110006
Device Lot Number(10) 86885
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/15/2021
Supplement Date Received by Manufacturer12/16/2022
Initial Report FDA Received Date01/04/2022
Supplement Report FDA Received Date12/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ROB00036
Patient SexUnknown
-
-